Interactive lab › Lock a laser to an atom and to a cavity
Interactive lab · 4/66 · Laser locking · ECDL
Lock a laser to an atom and to a cavity
Every cold-atom experiment needs lasers that stay within a fraction of a megahertz of an atomic line. An external-cavity diode laser, like the Cateye laser RCQT patented, is tuned onto a feature of a rubidium vapour cell or an optical cavity, and a feedback loop holds it there. Choose the reference and the detection, then disturb the laser.
What you are seeing
- Top: light through a rubidium cell. The broad dip is the Doppler-broadened absorption; the narrow peaks are the hyperfine lines and the crossovers half-way between them, revealed by a counter-propagating pump beam.
- Bottom: the error signal. Modulating the laser frequency and demodulating at the first harmonic gives the derivative of the spectrum; at the third harmonic it gives the third derivative, which flattens the sloping Doppler background so the zero crossing sits on the line itself.
- Pound–Drever–Hall uses phase-modulation sidebands and the light reflected from a high-finesse cavity: a far steeper slope, a capture range of ±Ω, and a loop that can be much faster.
- Press “Kick it”: within the capture range the loop pulls the laser back; beyond it the laser runs to another zero crossing or away.
The numbers in the model
| 87Rb D2 hyperfine lines from F = 2 | F′ = 3 · 2 · 1 at 0 · −266.65 · −423.60 MHz |
|---|---|
| Doppler width at room temperature (σ) | ≈ 220 MHz |
| Cavity in the model | free spectral range 1.5 GHz; linewidth = FSR / finesse |
| Loop | integrator; up to 700 Hz with a lock-in, 7 kHz with PDH |
| Dip heights | typical values, not calculated |
Go further
- The Cateye laser and its patents (film)
- Glossary: Pound–Drever–Hall locking
- Glossary: saturated absorption spectroscopy
- Lab: laser cooling, which needs these lasers