Build the error budget of a gravimeter

A measurement is only as good as its error budget: the list of everything that can shift or blur the result, each with a number. This is how RCQT designs its atom gravimeter for a target of a few µGal. Change the interferometer, the atoms, the site and the tricks such as k-reversal, and see which line dominates and how long you must average.

What you are seeing

  • Bars: standard uncertainty of each term in µGal (1 µGal ≈ one billionth of g), on a log scale. Blue are statistical and shrink with averaging; orange are systematic and do not.
  • Vibrations enter through the interferometer's response, which passes slow ground motion and filters fast motion; isolation and a seismometer correction cut it.
  • Hot atoms see a curved wavefront; atoms drifting sideways feel the Coriolis force; both are reduced by colder atoms, flatter optics and turning the instrument.
  • The Allan plot shows the statistical noise averaging down as 1/√τ until it meets the systematic floor: beyond that point, averaging longer does not help.

The numbers in the model

Effective wave numberk_eff = 4π/780 nm = 1.61 × 10⁷ m⁻¹
Phaseφ = k_eff g T²
Projection noise per shotδg = 1/(C √N k_eff T²)
Latitude used for Coriolis18.79° N (Chiang Mai)
1 µGal10⁻⁸ m/s² ≈ 1.02 × 10⁻⁹ g